Medical masks and respirators: Information for manufacturers

On this page

Regulatory requirements

Medical masks and respirators are regulated by Health Canada as Class I medical devices under the Medical Device Regulations (regulations).

Manufacturers, importers and distributors of Class I medical masks and respirators must have a medical device establishment licence (MDEL) before they can import or sell in Canada. There are exceptions for:

The responsibilities of an MDEL holder are described in section 45 of the regulations. MDEL holders must demonstrate to Health Canada that they have met the regulatory requirements. They must also have applicable documented procedures in place for the medical devices that they import or distribute.

A medical device licence is not required for Class I devices. However, manufacturers must comply with the following requirements set out in the regulations:

Non-medical masks and barrier face coverings are not regulated as medical devices by Health Canada.

Learn more about MDELs:

Regulating medical masks

Manufacturers of medical masks are expected to meet certain requirements before they import or sell in Canada.

Regulatory considerations

Masks are regulated as Class I medical devices if they are sold or represented for a medical purpose. For example, they help to prevent the spread of germs from the user to others or protect the user from germs.

Medical mask standards

Medical masks should conform with 1 or more medical mask standards, such as:

We may ask manufacturers to provide evidence when they claim that they are conforming to a standard or complying with the regulatory requirements.

Labelling

We consider it to be a claim when a mask is labelled that it conforms to a medical device standard. The evidence that we require depends on the standard to which the manufacturer claims performance. The standard defines the performance characteristics required for conformance.

Medical mask packaging should be labelled with the standard to which it claims conformance.

Reusable medical masks

Some medical masks may claim to be reusable up to a certain number of washes. Manufacturers of reusable medical masks should be able to:

Transparent medical masks

For information on the requirements for transparent medical masks, consult:

Regulating respirators

Manufacturers of respirators are expected to meet certain requirements before they import or sell in Canada.

Regulatory considerations

Respirators are regulated as Class I medical devices if they are sold or represented for a medical purpose. For example, they help to prevent the spread of germs from the user to others or protect the user from germs.

All surgical respirators are Class I medical devices.

Industrial respirators are not considered medical devices and are not regulated under the Food and Drugs Act if they:

Like respirators with medical claims, industrial respirators may be labelled with a filtration efficiency and be NIOSH-certified (for example, N95) as long as their intended industrial purpose is clear. These types of respirators are regulated under the Canada Consumer Product Safety Act.

Respirator certification

Certification organizations like the Canadian Standards Association (CSA) Group and the US National Institute for Occupational Safety and Health (NIOSH) evaluate, test and certify respirators. Certified respirators must pass minimum performance requirements, such as:

Respirators can be tested to Canadian standards or other international standards, such as:

Certified respirators will have a symbol of the certification body. Manufacturers should be able to provide evidence of certification for respirators that are labelled with a certification symbol.

For symbols, visit:

We may request evidence from respirator manufacturers that their respirators have been certified according to 1 or more of these standards. Alternatively, manufacturers can provide test reports that demonstrate their device is equivalent to a certified device.

In addition to standard respirator performance testing or certification, manufacturers of elastomeric and reusable respirators should:

We may request evidence from manufacturers of surgical respirators of fluid penetration and flammability testing.

Fluid penetration testing should be conducted in accordance with ASTM F1862/F1862M, as well as the respirator labelled with the barrier level achieved in this test. Flammability testing should be conducted in accordance with 16 CFR 1610.

It may be acceptable to test these characteristics using other standards. The manufacturer should provide a rationale if they use another standard.

Labelling

The label of a respirator should indicate the standard to which it conforms. For example, CSA-approved respirators and filters for reusable respirators must be labelled with the following:

Each respirator standard has its own labelling requirements. Manufacturers must conform with these requirements to remain in compliance with the standard.

Respirators that have not been certified by a certifying body should at least be labelled with the following:

Biocompatibility

Health Canada may ask manufacturers to provide evidence that their medical mask or respirator complies with the regulatory requirements to demonstrate the biocompatibility of the device. This usually involves providing evidence that the device is not:

Depending on the materials used in the device, it may be acceptable to demonstrate biocompatibility by comparing materials with an identical or very similar device that's available in Canada.

Nanomaterials or other coatings

Some medical masks and respirators may contain nanomaterials or coatings made from other substances. These coatings are usually intended to destroy or inhibit the growth of bacteria and viruses contacting the mask surface.

Examples of the materials used in these coatings include:

To date, there are no safety standards that directly address the use of nanomaterials or other coatings in respirators and medical masks. Instead of specific standards, we refer to the NIOSH-recommended exposure limit (REL) and the Occupational Safety and Health Administration (OSHA) permissible exposure limit (PEL).

Learn more about these limits:

Coating substances should be strongly fixed to the device to reduce the chance of particulates shedding and being accidentally inhaled.

The total mass of coating used in the mask should not be greater than the NIOSH's REL or OSHA's PEL limit, whichever is less. If the total mass is greater, manufacturers must be able to demonstrate that:

We may ask manufacturers to provide scientific evidence to:

For example, if:

For more information on regulatory considerations for nanomaterials or other coatings used with medical masks or respirators, consult: